Dual Incubator
Upper chamber Temperature Range
- 10°C to 60°C
- Accuracy: ±0.2°C
- Uniformity : ±1°C
lower Chamber Temperature Range
- 10°C to 60°C
- Accuracy: ±0.2°C
- Uniformity : ±1°C
Control System
Individual PLC based control system with fast Ethernet communication for each machine.
Individual 4.3” color touch screen display for each machine
Standard Features
This unit has double walled chamber, with PUF insulation.
Interior is made of Stainless Steel & Exterior is either made of Mild Steel finished with powder coating or Stainless Steel.
Inside full length observation glass door with secure gasket
The unit has outside metal door with gasket & lock.
The unit is provided with perforated Stainless Steel trays with adjustable height.
Interior illumination for working area
Forced air circulation for uniform temperature.
PU wheels for easy move ability
Sensor
PT-100 sensor used for measurement.
Temperature in degree Celsius with 0.1°C resolution.
Alarms
Audio and Visual alarms are generated for temperature variation, utility failure.
Safety
Unit is provided with safety devices for high and low temperature in case of malfunction.
Heating System
long lasting stainless steel tubular heaters with s.s. fins
Cooling System
CFC free, the cooling system consists of hermitically sealed compressor coupled with evaporation coil and condenser, safe guarded by time delay circuit.
Optional Upgrades
- SDMS Web based application for Alphatech Make Multiple Equipments
- Temperature Multiple channel Scanners.
- Mobile Alert System.
- Stand by cooling system for auto/Manual changeover.
- Door Access System.
- Remote Alarm system